← Back to Products
Contact for Inquiry
Class II B2 Biological Safety Cabinet
Total-exhaust containment for demanding applications. Class II B2 cabinets protect personnel, products and the environment while exhausting 100% of filtered air through a dedicated top connection. With no air recirculated into the work zone, the design is suited to procedures involving limited quantities of volatile chemicals or radionuclides when connected to a properly designed exhaust system.
0%air recirculation
100%filtered air exhaust
3-waypersonnel, product and environmental protection
Advantages
- Three-way protection for personnel, products and the surrounding environment.
- 100% filtered-air exhaust prevents air from being recirculated into the work zone.
- ULPA filtration reaches at least 99.9995% efficiency at 0.12µm.
- Stable inflow and vertical downflow help maintain the protective air barrier.
- A 200mm operating opening balances access with containment performance.
- The fully finished SS304 work area is durable and easy to clean.
- Integrated LED and UV lighting supports visibility and work-zone decontamination.
- The top exhaust outlet supports connection to a dedicated building exhaust system.
- Universal wheels with adjustable leveling feet simplify positioning and installation.
Model Comparison
| Model | YZ-BSC-1000IIB2 | YZ-BSC-1300IIB2 | YZ-BSC-1600IIB2 | YZ-BSC-1800IIB2 |
|---|---|---|---|---|
| Nominal Size | 1 meter (3') | 1.3 meter (4') | 1.6 meter (5') | 1.8 meter (6') |
| Filtration Efficiency | ULPA: ≥99.9995%, @0.12µm | |||
| Downflow Velocity | 0.35 m/s (68.88fpm) | |||
| Inflow Velocity | 0.55 m/s (108.24fpm) | |||
| Airflow System | 0% air recirculation, 100% air exhaust | |||
| Noise | 58~65 dB(A) | |||
| Vibration | ≤5 µm (center of table top) | |||
| Power Supply | AC220V, 1Φ, 50Hz | |||
| Opening Height | 200mm | |||
| Personnel Protection | A. Total colony in impaction sampler ≤10CFU./time B. Total colony in slot sampler ≤5CFU./time | |||
| Product Protection | Total colony in culture dish ≤5CFU./time | |||
| Cross-contamination Protection | Total colony in culture dish ≤2CFU./time | |||
| Max Consumption (with spare socket) | 1.65 kVA | 1.70 kVA | 1.75 kVA | 1.75 kVA |
| Rated Power (without spare socket) | 0.33 kVA | 0.38 kVA | 0.43 kVA | 0.43 kVA |
| Weight | 270 kg | 310 kg | 370 kg | 410 kg |
| Work Dimension (W1×D1×H1) | 1004×630×630 mm (39.5″×24.8″×24.8″) | 1304×630×630 mm (51.3″×24.8″×24.8″) | 1604×630×630 mm (63.1″×24.8″×24.8″) | 1804×630×630 mm (71.0″×24.8″×24.8″) |
| Overall Dimension (W×D×H) | 1200×795×2050 mm (47.2″×31.3″×80.7″) | 1500×795×2050 mm (59.1″×31.3″×80.7″) | 1800×795×2050 mm (70.9″×31.3″×80.7″) | 2000×795×2050 mm (78.7″×31.3″×80.7″) |
| Size and Qty. of Light | LED 24.5 W × 1 | LED 31 W × 1 | LED 36 W × 1 | LED 36 W × 1 |
| Size and Qty. of UV Light | 18 W × 1 | 30 W × 1 | 36 W × 1 | 36 W × 1 |
| Lumin. | ≥900 Lx | |||
| Cabinet Materials | High-grade steel and lacquered in ivory | |||
| Working Area Materials | SS304 fully finished | |||
| Air Direction | Top out | |||